DEFINITION OF TERMS RELATED TO HUMIDITY:
There are many terms
associated with humidity which can lead to confusion.
Relative humidity
Relative humidity is defined
as the ratio of the water vapor pressure to the saturation vapor pressure
(over water) at the temperature of the gas.
RH = Pw/Pws * 100
Water vapor
Water vapor can be treated as
a gas. At a particular temperature air for example can only hold so much
water vapor. The higher the temperature the more water vapor it can hold.
When saturated the relative humidity would be 100 % RH, so relative humidity
describes how close to saturation the air is. It is important to remember
that in a process with a high relative humidity a small drop in temperature
will cause the humidity to rise and the environment to saturate. Rapid
temperature changes in a an environmental chamber for example could also
cause condensation.
Dewpoint
Is the temperature at which
air becomes saturated with water and begins to condense - forming a dew.
Therefore at 100 % RH the ambient or process temperature equals the dewpoint
temperature. The more negative the dewpoint temperature is from the ambient
temperature the less the risk of condensation and the drier the gas or air
stream.
Wet bulb
Traditionally this was the
temperature indicated by a thermometer whose bulb is wrapped in a wet
sheath. The wet bulb temperature and the dry bulb temperature (i.e air
temperature) would then be used to calculate relative humidity or dewpoint.
Alternatively charts or tables can be used.
Mixing ratio
Is often used in drying
applications and is the ratio of the mass of water vapor to the mass of dry
air with which it is associated. Vaisala products give this output as grams
of water per kilogram of dry air (g/kg).
Absolute humidity
This is often confused with
mixing ratio. It is the ratio of the mass of water vapor to the unit volume
of moist air. Vaisala products give this output as grams of water per cubic
metre of air (g/m3).
ELECTRONIC RH SENSORS
DEFINITION OF TERMS RELATED
TO ELECTRONIC RH SENSORS:
When looking at the
specification of an electronic RH sensor the following should be taken into
account
Accuracy
Accuracy is a much abused term
in humidity measurement because it is difficult to define and many
unrealistic claims are made. However it is still regarded as a major factor
when choosing an instrument. Of equal importance are other factors such as
stability, hysteresis, temperature dependence and chemical tolerance. Also
see the section on calibration. The majority of Vaisala products are
compared and adjusted against factory standards during manufacture. These
factory standards are traceable to various national standards. Some Vaisala
products are automatically supplied with Vaisala calibration certificates.
Linearity
This is a measure of the
offset from a specified straight line joining the end points of the working
range. In the case of Vaisala sensors the accuracy specification includes
the linearity and hysteresis errors.
Hysteresis
This is the difference in
output in %RH when the humidity is increased from a set point by a specified
amount then decreased by the same amount (e.g from 10 %RH to 75 %RH then
back to 10 %RH, over a one hour period). Vaisala includes this error in the
accuracy figure.
Repeatability
Describes the ability of a
sensor to repeat its performance in a number of tests.
Stability
Gives an indication of drift
over a period of time. Typically the HUMICAP sensor has stability of
±0.5%RH/year in normal air conditions, which makes it the most stable
humidity sensor available today.
Temperature dependence
Shows the effect of
temperature change on the RH output value. Over a large temperature range
this change can be substantial. In Vaisala products the temperature
compensation built in.
Response time
Is the speed of response of
the sensor when subjected to a step change in humidity. This varies
depending on the temperature, air flow and type of filter fitted. A typical
response time with Vaisala products is 15sec.
Vaisala RH meters employ the
following types of sensors:
Relative Humidity (RH)
Vaisala produced the World’s
first thin film capacitive humidity sensor in 1973. Described as the HUMICAP
the technology of this sensor has been continuously improved over the years.
The most important features of the sensor are the operating range (relative
humidity and temperature), stability, accuracy, repeatability, hysteresis,
chemical tolerance and speed of response.
The principle of operation of
the sensor is based upon a change in capacitance as a polymer film absorbs
and desorbs water vapor. The input stage of the transmitter electronics must
also be capable of measuring this small non linear capacitance change with
good accuracy and repeatability.


Temperature (T)
Many Vaisala products also
have a temperature sensor built in because it is often important to know the
temperature of the environment at the point where the humidity is being
measured. Vaisala instruments use platinum resistance Pt100 or Pt1000
sensors to measure temperature. This type of sensor is a precision resistor
which increases in resistance with temperature. The platinum as a material
does not change over time, so the change in resistance is very stable and
repeatable.
VARIABLES IN ELECTRONIC RH
MEASUREMENT:
Temperature differences
It is important that the
humidity sensor is at the same temperature as the process or environment
being measured. At 20C and 50 %RH a +/-1C difference between the sensor and
zone of measurement will cause an error of +/-3 %RH. At 90 %RH the error is
+/-6 %RH. At high humidities a temperature difference of a few degrees can
cause water to condense on the sensor. In an unventilated space it may take
hours to dry out. Vaisala humidity sensors will start to function normally
as soon as the water has evaporated but if the water is contaminated the
life of the sensor may be shortened and calibration may change.
Other influences
Care must be taken to avoid
hot or cold spots that are not representative of the environment being
measured. A sensor located near a door, room humidifier, heater or air
conditioning inlet duct for example would be subject to rapid humidity or
temperature changes and may appear to be unstable. Avoid also sample flows
where the gas temperature can drop below the dewpoint temperature.
Chemical tolerance
Certain chemicals in the
atmosphere surrounding the probe head could also attack or contaminate the
sensor. Vaisala sensors have good chemical tolerance and we have a
comprehensive list of the effect of various concentrations on sensor
performance.If you have any queries about a particular chemical please
contact us. It is important to know if the sensor head could be subjected to
any periodic sterilization routines.
ELECTRONIC RH CALIBRATION
In case there is no high
accuracy reference meter available, the Vaisala products with a ± 1%RH
accuracy is recommended to send back to the Vaisala Lab for recalibration.
With ± 2%RH and ± 3%RH accurate products also the salt solution
calibration(see below) can be used.
Laboratory calibration
Relative humidities can be
generated in specially adapted chambers using various saturation techniques
and compared against transfer standard humidity or dewpoint probes. However
the cost is significant and can only be justified where there is a need for
regular humidity calibrations. Saturated salt calibration at two points is
recommended for the majority of Vaisala probes. Using this method standard
value relative humidity environments are generated in small containers
containing selected aqueous saturated salt solutions. The Vaisala HMK11 uses
this principle. A saturated solution of Lithium Chloride (LiCl) for example
will generate a constant humidity of 11.3 %RH and Sodium Chloride (NaCl)
will generate a humidity of 75.5 %RH (both at 20°C). See table below.
Under ideal conditions the
accuracy is better than ± 1%RH. However due to possible temperature
variations and other error sources the accuracy is reduced to ± 2.5 %RH.
Other salt values are also available. For further information on saturated
salts please refer to the American ASTM standard E104-85, the German
standard DIN50008 and the work by L. Greenspan, published in the Journal of
Research by the National Bureau of Standards (now NIST) Vol 81A.
Temperature stability
and equilibrium
When making comparisons
between a calibrated meter and a fixed probe for example it is important
that the two probe heads are as close together as possible and they are at
the same temperature. For example at 50 %RH and 20°C a 1°C difference
between the 2 heads would mean a difference in RH of 3 %RH. At 90 %RH the
corresponding error is 6 %RH. Rapid changes in temperature will also make a
calibration check difficult. This consideration is particularly important
when a meter is being used as a transfer standard to check a number of fixed
installation probes in different ambient conditions. The temperature of the
meter sensor head, hence the temperature readout, must be allowed to
stabilise before a comparison reading is taken.
In applications where a probe
is to be operated in extremes of temperature or over a wide temperature
range the temperature dependence of the sensor and electronics must also be
taken into account. For example a temperature dependence of ± 0.04 %RH/°C
means that over an operating temperature range of 50°C the change in RH
could be up to 2 %RH.
Comparisons between
instruments and other humidity measurement techniques:
RH (relative humidity)
measurement can be accomplished by several means. A sling psychrometer can
be used to measure RH. This is done using a wet and a dry thermometer; the
RH is proportional to the temperature of the wet bulb when compared to the
temperature of the dry thermometer. This system is tried and true but not as
accurate as digital and synthetic hair hygrometers.
Synthetic Hair Hygrometers:
Synthetic hair hygrometers are
exactly that—a moisture sensitive "hair" that contracts or expands depending
on the amount of water in the air. This contraction and expansion can easily
be measured on a dial marked with the % RH. Synthetic hair hygrometers are
accurate, but only in a limited range. If your application involves small
fluctuations in RH%, then these are economical and effective devices. We do
not recommend these as portable units, because the reaction time is slow,
and moving rapidly through changing environments will affect the calibration
and accuracy of the units.
Problems often occur when
using different techniques to compare and calibrate humidity probes. For
example whirling hygrometers often give a higher reading when compared to a
fixed probe. Only the thermometers in a whirling hygrometer can be
calibrated and errors are often caused in such instruments through poor
maintenance, incorrect usage and temperature differences. Errors can easily
be introduced when using conversion tables as the relationships between
humidity variables are complex. In cases of dispute the calibration of the
instruments involved and their traceability should be carefully checked. For
example if two meters have a specified RH accuracy of ±2 %RH and ±3 %RH
respectively the difference in reading could be up to 5 %RH and both meters
would theoretically be in calibration.
Frequency of calibration
This depends on the relative
humidity, temperature and other environmental factors. As a general guide a
calibration period of 1 year is recommended for temperatures up to 35°C and humidities up to 70 %RH. Above these values the humidity should initially be
checked at six monthly intervals.
Humidity calibration
certificates
An individual certificate
showing two or more comparison points will give an indication of the
accuracy and linearity of a humidity product at the time of calibration. A
NIST calibration certificate can be provided with an additional cost.